    [S.l.] : American Institute of Physics (AIP)
    Review of Scientific Instruments 73 (2002), S. 3122-3124 
    ISSN: 1089-7623
    Source: AIP Digital Archive
    Topics: Physics , Electrical Engineering, Measurement and Control Technology
    Notes: We report on the continuous measurement of fluorescence lifetimes at low light levels. Fluorescence photons following pulsed excitation generate a pulse sequence with exponentially distributed amplitudes and interphoton times at the output of a time-to-amplitude converter. This sequence is turned into a continuous step function and is time averaged with an adjustable bandwidth. For a single-exponential decay, our approach yields identical results as would be obtained from fitting fluorescence decays, while being a real-time technique. The proposed technique performs especially well at low count rates. We demonstrate the applicability of the method at the example of confocal fluorescence lifetime imaging of single molecules. © 2002 American Institute of Physics.