    Notes: Tubings extruded from one batch of polyvinylchloride resin at five temperatures ranging from 110 to 205°C were tested for thrombogenicity in the saphenous vein and aorta of sheep. There was more thrombus in both the vein and aorta on tubings extruded at lower temperatures. The surface areas of renal infarction were smaller when lower temperature tubings were inserted into the aorta. Scanning electron microscopy showed that lower extrusion temperatures caused greater roughness on the tubings. This greater roughness was confirmed by roughness measurements. It is suggested that the association which has been demonstrated between roughness and thrombus formation might be due to the degree of adhesion of the thrombus to the surface rather than to differences in thrombogenicity.